heto- holten model safe 2010-1.2 class II safety
cabinet
Bench-top
working place requiring safe area with respect to operator, product
and environment.
Specifications
The front window is sloping sash type with electric host unit.
It is sealed completely when the UV light is turned on.
Working table is made of divided stainless steel for easy cleaning
and autoclaving.
The cabinet is equiped with two HEPA filters in the top, one
main fitler an done exhaust fitler. 75 % of the air volume is
recirculated in the working place, 25 % is exhausted from exhaust
fitler.
HEPA filter efficiency is 99,99 % for particles bigger than
0,3 µm
The cabinet has microprossesor controller with safety alarms
and LCD display. The clock for presetting automatic start-up,
reduced speed, UV-timer, indication of filter change and alarm
for any malfunction is easily fallowed from LCD-display.
All air cannels that can be contaminated are surrounded by
negative pressure
Applications
It provides safe area with autoclavable divided surfaces,
UV-light and heat sterilization with its gas valve to protect operator,
product and environment.